Perusahaan ini terdaftar sebagai Free Member. Hindari melakukan pembayaran sebelum bertemu penjual atau melihat barang secara langsung. COD (Cash On Delivery) atau bertemu langsung dengan penjual merupakan metode transaksi aman yang kami sarankan.
DR900 Multiparameter Portable Colorimeter Hach |
|
Specifications |
|
Battery Life |
6 months ( typical ) @ 5 readings a day / 5 days / week without backlight** |
** Backlight usage will decrease battery life. |
|
Battery Requirements |
4, AA size alkali cells |
Data Logger |
500 measured values ( Result, Date, Time, Sample ID, User ID ) |
Detector |
Silicon photodiode |
Dimensions |
H 231 mm x W 96 mm x D 48 mm |
Display |
Graphical display 240 x 160 pixels ( Backlit ) |
Enclosure Rating |
IP67 ( vial cover closed ) |
Includes |
DR 900 Colorimeter, two 1-inch glass sample cells marked at 10, 20 and 25 mL, two 1 cm plastic sample cells, 1 x 16-mm COD/Test 'N Tube™ adapter, 4 AA alkaline batteries, printed multilingual instrument manual, instrument and procedure manuals on CD, USB Mini to USB Cable, European CE mark. |
Interface |
USB type Mini IP67 |
Manual Languages |
English, French, German, Italian, Spanish, Portuguese, Bulgarian, Chinese, Czech, Danish, Dutch, Finnish, Greek, Hungarian, Japanese, Korean, Polish, Romanian, Russian, Slovenian, Swedish, Turkish |
Operating Conditions |
10 - 40 °C ( 50 - 104 °F ), max. 80 % RH ( non- condensing ) |
Operating Humidity |
Max 90 % RH ( non-condensing ) |
Operating Mode |
Transmittance ( % ), Absorbance and Concentration |
Operating Temperature |
10 - 40 °C |
Optical System |
0 / 180 deg transmittance |
Photometric Accuracy |
± 0.005 Abs @1.0 ABS Nominal |
Photometric Linearity |
± 0.002 Abs ( 0 - 1 Abs ) |
Photometric Measuring Range |
0 - 2 Abs |
Power Supply |
4 x AA size alkaline cells |
4x NiMH rechargeable Battery* |
|
( *External charger required ) |
|
Reproducibility |
± 0.005 Abs ( 0 - 1 A ) |
Sample Cell Compatibility |
1 inch round or 16 mm round ( with adapter ) |
Spectral Bandwidth |
15 nm filter bandwidth |
Storage Conditions |
-30 to 60 °C ( C68-30 - 140 °F ), max. 80 % RH ( non-condensing ) |
Supported Chemistry |
Hach tests |
Temperature Range |
0 - 50 °C ( 32 - 122 °F ) |
User Interface Languages |
English, French, German, Italian, Spanish, Portuguese, Bulgarian, Chinese, Czech, Danish, Dutch, Finnish, Greek, Hungarian, Japanese, Korean, Polish, Romanian, Russian, Slovenian, Swedish, Turkish |
User Programs |
Custom programming 10 |
Wavelength Accuracy |
± 1 nm ( fixed, varies with model ) |
Wavelength Calibration |
N/A |
Wavelength Range |
420 nm, 520 nm, 560 nm, 610 nm |
Wavelength Selection |
Automatic |
Weight |