Perusahaan ini terdaftar sebagai Free Member. Hindari melakukan pembayaran sebelum bertemu penjual atau melihat barang secara langsung. COD (Cash On Delivery) atau bertemu langsung dengan penjual merupakan metode transaksi aman yang kami sarankan.
Thermo Scientific GENESYS 50 UV-Visible & GENESYS 40 Vis Spectrophotometer |
||
Specifications |
||
Product Size |
GENESYS 50 UV-Vis Spectrophotometer |
GENESYS 40 UV-Vis Spectrophotometer |
Part Number |
840-298000 |
840-297000 |
Baseline Flatness |
±0.002 A |
±0.005 A |
Compartments |
Accessible from top, front or side |
|
Able to accommodate cells up to 100 mm cells |
||
Available accessory for test-tubes up to 25 mm diameter and 150 mm height |
||
Removable, washable sample compartment liner with magnetic placement and hold-down |
||
Connections |
Single USB-A supports flash memory devices for method and data storage |
|
Duplex USB-A on side supports connection to a Windows™ computer running optional remote control software, keyboard, mouse |
||
Export data to network or PC via USB, Ethernet, or Wi-Fi |
||
Print via USB, Ethernet or Wi-Fi |
||
Description |
GENESYS 50 UV-Vis Spectrophotometer |
GENESYS 40 Visible Spectrophotometer |
Detector Type |
Dual Silicon Photodiodes |
|
Dimensions ( L x W ) |
35.5 x 38.5 x 19.5 cm |
|
Display |
7-inch color touchscreen, fixed, high definition, 800 × 1280 pixels |
|
Drift |
( At 500 nm after 1 hour warmup ) |
( At 500 nm after 1 hour warmup ) |
Electrical Requirements |
External AC to DC converter. Voltage and Frequency ( Hz ) selected automatically, 100–240 volts, 50–60 Hz. |
|
Keypad |
Touchscreen |
|
Lamp |
Xenon Flash Lamp ( >5 years typical, 3 years guaranteed ) |
Tungsten-Halogen ( >1000 hours ) |
Noise |
( RMS at 500 nm 60 consecutive measurements ) |
( RMS at 500 nm 60 consecutive measurements ) |
≤0.00020 A at 0 A at 260 and 500 nm |
||
≤0.00030 A at 1 A at 260 and 500 nm |
||
≤0.00040 A at 2 A at 260 and 500 nm |
||
Optical Design |
Dual beam |
|
Photometric Accuracy Instrument |
±0.002 A at 0.5 A |
|
±0.004 A at 1.0 A |
||
±0.008 A at 2.0 A |
||
Photometric Display |
-3 A to +5 A |
|
Photometric Range |
–2 A to +3.5 A |
–3 A to +3.5 A |
Photometric Repeatability |
±0.001 A at 1 A measured at 1.0 A at 546 nm |
|
Printer |
Snap-on printer available |
|
Spectral Bandwidth |
2 nm |
5 nm |
Stray Light |
< 1.0 % T 198 nm ( KCl ) , |
|
Wavelength Accuracy |
±0.5 nm |
|
Wavelength Data Interval |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
|
Wavelength Range |
190 nm – 1100 nm |
325 nm – 1100 nm |
Wavelength Repeatability |
< ±0.2 nm |
|
Wavelength Scan Speed |
Slow, medium and fast - up to 1600 nm/min |
Automatic – up to 1800 nm/min |
Weight ( Metric ) |
7.5 kg |