Perusahaan ini terdaftar sebagai Free Member. Hindari melakukan pembayaran sebelum bertemu penjual atau melihat barang secara langsung. COD (Cash On Delivery) atau bertemu langsung dengan penjual merupakan metode transaksi aman yang kami sarankan.
Constant Climate Chamber P 240 & KBF P 720 Binder |
||||
Designation |
KBFP240-230V¹ |
KBFP240UL-240V¹ |
KBFP720-230V¹ |
KBFP720UL-240V¹ |
Option model |
Standard |
Standard |
Standard |
Standard |
Order number |
9020-0328 |
9020-0329 |
9020-0330 |
9020-0331 |
Measures |
||||
Interior volume |
247 L |
247 L |
700 L |
700 L |
Net weight of the unit ( empty ) |
213 kg |
213 kg |
374 kg |
374 kg |
permitted load |
100 kg |
100 kg |
150 kg |
150 kg |
Load per rack |
30 kg |
30 kg |
45 kg |
45 kg |
Doors |
||||
Inner doors |
1 |
1 |
2 |
2 |
Unit doors |
1 |
1 |
2 |
2 |
Internal Dimensions |
||||
Width |
650 mm |
650 mm |
973 mm |
973 mm |
Height |
785 mm |
785 mm |
1250 mm |
1250 mm |
Depth |
485 mm |
485 mm |
576 mm |
576 mm |
Outer dimensions |
||||
Width net |
925 mm |
925 mm |
1250 mm |
1250 mm |
Height net |
1460 mm |
1460 mm |
1925 mm |
1925 mm |
Depth net |
800 mm |
800 mm |
890 mm |
890 mm |
Wall clearance sidewise |
100 mm |
100 mm |
200 mm |
200 mm |
Wall clearance back |
100 mm |
100 mm |
100 mm |
100 mm |
Environment-specific data |
||||
Energy consumption at 40 °C & 75 % RH |
760 Wh/h |
760 Wh/h |
1850 Wh/h |
1850 Wh/h |
Sound-pressure level |
53 dB ( A ) |
53 dB ( A ) |
53 dB ( A ) |
53 dB ( A ) |
Fixtures |
||||
Number of shelves ( std./max.) |
2 - 7 |
2 - 7 |
3 - 12 |
3 - 12 |
Number of illumination cassettes ( std./max.) |
1 |
1 |
1 |
1 |
Fixtures |
||||
Number of illumination cassettes ( std./max.) |
2 - 2 |
2 - 2 |
3 - 3 |
3 - 3 |
Performance Data Temperature |
||||
Temp.range without illumination cassettes |
0…70 °C |
0…70 °C |
0…70 °C |
0…70 °C |
Temp.range with 100 % illumination |
10…60 °C |
10…60 °C |
10…60 °C |
10…60 °C |
Max. heat compensation at 40 °C |
400 W |
400 W |
1000 W |
1000 W |
Performance Data Climate |
||||
Temp. range with 100 % illumination |
10…60 °C |
10…60 °C |
20…60 °C |
20…60 °C |
Temp.range without illumination cassettes |
10…70 °C |
10…70 °C |
10…70 °C |
10…70 °C |
Temp. variation with illumination at 40 °C & 75 % RH |
±0.6 K |
±0.6 K |
±1.2 K |
±1.2 K |
Temp. variation with illumination at 25 °C & 60 % RH |
±0.6 K |
0.6 K |
±1.2 K |
±1.2 K |
Temp. fluctuation with illumination at 25 °C & 60 % RH |
±0.2 K |
0.2 K |
±0.2 K |
±0.2 K |
Temp. fluctuation with illumination at 40 °C & 75 % RH |
±0.2 K |
0.2 K |
±0.2 K |
±0.2 K |
Humidity range with illumination cassettes, with illumination |
10…75 % RH |
10…75 % RH |
10…75 % RH |
10…75 % RH |
Humidity range without illumination |
10…80 % RH |
10…80 % RH |
10…80 % RH |
10…80 % RH |
Humidity fluctuation with illumination at 40 °C & 75 % RH |
±2 % RH |
±2 % RH |
±2 % RH |
±2 % RH |
Humidity fluctuation with illumination at 25 °C & 60 % RH |
±1.5 % RH |
±1.5 % RH |
±2 % RH |
±2 % RH |
Humidty recovery time after 30 s door open at 40 °C & 75 % RH with illumination |
6 min |
6 min |
5 min |
5 min |
Humidty recovery time after 30 s door open at 25 °C & 60 % RH with illumination |
4 min |
4 min |
4 min |
4 min |
Light data per Illumination Cassettes |
||||
ICH compliant illumination for photo stability testing [ lx ]* |
7500 |
7500 |
9000 |
9000 |
ICH compliant illumination for photo stability testing [ W/m² ]* |
1.1 |
1.1 |
1.5 |
1.5 |
Electrical data |
||||
Rated Voltage |
200…230 V |
200…240 V |
200…230 V |
200…240 V |
Power frequency |
50/60 Hz |
50/60 Hz |
50/60 Hz |
50/60 Hz |
Nominal power |
0 |
2.4 kW |
3.5 kW |
3.5 kW |
Unit fuse |
16 A |
16 A |
16 A |
16 A |
Phase ( Nominal voltage ) |
1 Phase |
1 Phase |
1 Phase |
1 Phase |