Chlorine DPD Tablet Colorimeter Test Kit Lamotte |
|
Product Features |
|
Range |
0 to 4.0 ppm Chlorine ( may be extended by dilution ) |
Method |
In the absence of Iodide, Free Available Chlorine reacts instantly with buffered diethyl-p-phenylenediamine indicators ( DPD ) to produce red color in proportion to amount of chlorine present |
Subsequent addition of potassium iodide produces rapid color response from the combined forms of chlorine ( chloramines ) |
|
Handling/Preservation |
Chlorine in aqueous solutions, particularly weak solutions, is not stable |
Exposure to sunlight or agitation will accelerate reduction of chlorine |
|
Fill sample containers to top and cap tightly |
|
Analyze samples as soon as possible after collection |
|
Interferences |
The only interfering substance likely to be encountered is oxidized manganese |
The extent of this interference can be determined by treating samples with sodium arsenite to destroy present chlorine present, so that amount of interference can be measured |
|
Calibration |
The single test colorimeter is precalibrated |
In order to comply with NPDWR or NPDES reporting regulations, calibration should be checked periodically by using reference standards, including a 0 mg/L blank and 0.3, 1.0, and 3.5 mg/L chlorine |
|
Consult local regulatory agency to determine best practices and procedures to prepare necessary standards |
|
Applications |
|
Drinking water supplies |
|
Water distribution systems |
|
Swimming pools |
|
Spas |
|
Sewage |
|
Chlorinated waste waters |
|
Processed waters |
|
Sanitizing solutions |