Ovens YCO-010 & YCO-N01 Gemmy |
|
Specification |
|
Product |
Hot Air Oven |
Model |
YCO-010 |
Capacity |
16, 40, 57, 75, 90, 110, 150, 200 L |
Control system |
Microprocessor system |
Temp control |
PID |
Temp setting |
Digital |
Temp readout |
Digital |
Air convection |
Forced convection by fan |
Temp uniformity |
±3.0 ºC at 180 ºC |
Temp stability |
±0.5 ºC at 180 ºC |
Over-temp alarm & shutout |
Yes, 10 ºC to the set point |
Timer |
Digital, 9 hr. 59 min. + HOLD |
Temperature range |
Ambient + 5 ºC to 250 ºC |
Standard accessory |
2 S/S mesh shelves, adjustable. |
Key Features |
|
1 Temperature uniformity & Stability are approximated which may be varied under the different operation environment |
|
2 The above specifications are subject to change without advanced notice. |
|
GEMMY Hot Air Oven Sterilizer is designed for general purpose in applications in variety of baking, drying, conditioning, preheating, curing for dry content analysis, chemical resistance studies, glassware drying, dry sterilizer etc. |
|
All equipment features double wall construction for durability and good heat retention. |
|
Heavy-gauge glass fiber insulation sustains to prevent the heat loss and maximize energized efficiency. |
|
Timer is working only when the actual temperature arrived 2 ºC below set point. |
|
LED digital set and display for Power, Heating, PV, SV and Alarm. |
|
Adjustable fresh air inlet port at bottom and top provides control of the amount of convective air. |
|
Quality surface treated by power coating ensures scratch free and corrosion free. |
|
Safety devices : |
|
Over temperature alarm and shutout device to 10 ºC above the set point. |
|
Independent safety switch to shut out power protecting oven from excessive heating at 220 ºC if primary control fails |
|
Optional for current leakage breaker with O.C. |
|
Audible & visible alarm indicated by indicator. |
|
Timer stop counting when PV is 5 ºC below SV after stabilization if door is opened, the memory will store the past time and start to count when the PV value returns to 2 ºC below SV. |